Abstract

Multicolor single molecule localization-based super-resolution microscopy (SMLM) approaches are challenged by channel crosstalk and errors in multi-channel registration. We recently introduced a spectral demixing-based variant of direct stochastic optical reconstruction microscopy (SD-dSTORM) to perform multicolor SMLM with minimal color crosstalk. Here, we demonstrate that the spectral demixing procedure is inherently free of errors in multicolor registration and therefore does not require multicolor channel alignment. Furthermore, spectral demixing significantly reduces single molecule noise and is applicable to astigmatism-based 3D multicolor imaging achieving 25 nm lateral and 66 nm axial resolution on cellular nanostructures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.