Abstract
Results from an instrument-level spectral characterization of the 21-channel High Resolution Dynamics Limb Sounder (HIRDLS) flight instrument will be presented. These data were obtained during the prelaunch calibration of HIRDLS at the University of Oxford (fall 2002). A monochromator, equipped with a controllable diffraction grating, was used to produce monochromatic light for these tests. The monochromator was housed, along with HIRDLS, in a large vacuum chamber. The monochromator was also equipped with a polarizer, which allowed data to be acquired at known polarizations for each channel. A calibration detector, with a flat spectral response, was used to measure the output from the monochromator. This paper will document the analysis procedures used to obtain a measured instrument spectral response for each channel, along with the associated error analyses for these measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Geoscience and Remote Sensing
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.