Abstract

The Cross‐track Infrared Sounder (CrIS) radiometric accuracy depends upon accurate frequency calibration. Here we present both the prelaunch calibration of the sensor and the minor modifications needed to this calibration post launch. Particular emphasis is given to ensuring that all nine detectors on each of the three CrIS focal planes are on a common frequency scale with accurate off‐axis apodization corrections. Radiances from the current operational algorithm have a frequency calibration that is stable to 2 ppm, although this work suggests that the CrIS instrument is capable of a frequency calibration stability of better than 1 ppm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.