Abstract
The mechanisms responsible for the maintenance and termination of ventricular fibrillation (VF) are poorly understood. The aim of this study was to compare the spectral characteristics of the electrical signal during sustained and non-sustained VF in patients with an implantable cardioverter-defibrillator. The study included 51 patients who had had at least one episode of sustained VF (i.e., duration >5 s and requiring shock administration) and non-sustained VF (i.e., duration >3 s and spontaneously terminated) that were recorded by the device set in a unipolar configuration. Spectral analysis of the first 3 s of each episode was performed. The dominant frequency was higher in sustained VF (4.6+/-0.7 Hz) than in non-sustained VF (4.3+/-0.6 Hz; P=.01), while the other parameters were similar. Although the spectral characteristics of sustained and non-sustained VF were similar, differences were observed during the first 3 s that could be used in algorithms for the early detection of non-sustained VF.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.