Abstract
In this study we appraised the accuracy of the Finometer in spectral analysis of systolic (SAP) and diastolic (DAP) arterial pressure. In 24 patients the blood pressure signal was simultaneously recorded invasively and by the Finometer and the power of SAP and DAP in the VLF (0.01/spl divide/0.04 Hz), LF (0.04/spl divide/0.15 Hz), and HF (0.15/spl divide/0.45 Hz) bands computed. To investigate the relationship between invasive and Finometer measurements, the percentage error was computed and bivariate spectral analysis performed. The error for SAP analysis, was (%): 139, 76, -32, 75 for respectively VLF, LF, HF and total power; DAP errors were much smaller. SAP variability is greatly enhanced by the Finometer, mainly due to the increase of power in the VLF and LF bands which overrides the reduction in the HF band. Noise and nonlinearities account for < 20% of the Finometer power changes, indicating that the differences are mainly due to amplification or attenuation effects.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.