Abstract
Both the DPD and amperometric titration methods for measuring free available chlorine (FAC) are subject to interference from monochloramine and organic chloramines. The measured FAC concentration in the presence of chloramines is often significantly larger than the true FAC residual. Interference in the DPD method can be reduced by rapid titrations. Interference in the amperometric titration method can be minimized by maintaining a 200‐mV positive potential on the platinum electrode.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.