Abstract
Abstract A miniature Centrifugal Fast Analyzer has been modified for fluorescence and light-scatter measurements by using several rotors developed for this purpose. The modified system has been used to evaluate the feasibility of adapting specific protein analyses, such as IgG, IgA, IgM, C'3 complement component, and α-1-antitrypsin, to the Centrifugal Fast Analyzer. A study of reaction conditions has revealed that the addition of polyethylene glycol 6000 (Carbowax) to the dilution medium increases the rate of reaction and allows apparent equilibrium to be achieved in less than 60 s. Furthermore, scatter intensity is enhanced. This system can be used to make rapid immunoglobulin measurements with only microliter volumes of antibody (2-7 µl), without the need of sample blanks. Determination of antigen excess by a method that involves dynamic injection of antibody is also discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.