Abstract

A very specific, selective, simple, and inexpensive procedure was developed for the speciation of CrVI and CrIII. This method is based on the quantitative extraction of chromate and CrIII (previously oxidized to CrVI) as a tetrabutylammonium-chromate ion-pair in methyl isobutyl ketone (MIBK), and then back extraction and preconcentration with an acidic diphenylcarbazide (DPC) solution. Back extraction was applied to achieve further preconcentration by a final factor of 20. The CrVI-DPC complex was determined in back-extract by a spectrophotometer at 548 nm. Under these extraction conditions, most of the probable concomitant cations and anions remained in the first inorganic phase. The calibration curve was linear up to 0.14 microg L(-1) of CrVI with a detection limit of 2.22 ng L(-1). The developed procedure was found to be suitable for the determination of the CrVI and CrIII species in various natural water samples with a relative standard deviation of better than 1.6%. The method was successfully applied to the speciation of chromium in spiked natural water samples, and also samples of effluent from a leather treatment plant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.