Abstract

In this paper, we use a human head and shoulder model to extend the analysis of SAR in the human head due to radiation caused by UHF RFID readers. We use an FEM human head and shoulder model with a voxel size of 8 mm for the tetrahedral and 25 distinct tissues to analyse the peak one-voxel SAR, spatial-peak 1 g cube of tissue SAR, spatial-peak 10 g cube of tissue SAR and the average SAR in the human head. We study the SAR in the human head at 1 W radiated power of a 7.4 dB RFID reader patch antenna at distances of 10 cm, 100 cm and 1,000 cm from the human head. We show that an RFID reader at 10 cm from the human head presents a SAR above 1.6 W/kg for both the spatial-peak 1 g and 10 g cube of tissue, the maximum value allowed by the FCC in the USA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.