Abstract

The specific absorption rate (SAR) is an important safety criterion, limiting many MR protocols with respect to the achievable contrast and scan duration. Parallel transmission enables control of the radiofrequency field in space and time and hence allows for SAR management. However, a trade-off exists between radiofrequency pulse performance and SAR reduction. To overcome this problem, in this work, parallel transmit radiofrequency pulses are adapted to the position in sampling k-space. In the central k-space, highly homogeneous but SAR-intensive radiofrequency shim settings are used to achieve optimal performance and contrast. In the outer k-space, the homogeneity requirement is relaxed to reduce the average SAR of the scan. The approach was experimentally verified on phantoms and volunteers using field echo and spin echo sequences. A reduction of the SAR by 25-50% was achieved without compromising image quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.