Abstract

Selenium is one of the trace and essential elements for good health but required only in a very narrow range. Hence, determination of selenium in trace level is important. In this work, cloud point extraction (CPE) with non-ionic surfactant Triton X-114 and the fluorometric ligand, 2,3-diaminonaphtalene (DAN) were used for extraction of trace amounts of organic and inorganic selenium species prior to their determination by spectrofluorometry. CPE parameters affecting complexation and phase separation were optimised. The limit of detection calculated by using nine replicate measurements of 0.020mg/L Se solution after complexing with DAN and 10-fold CPE preconcentration was 2.1μg/L. Accuracy of the method was checked using EnviroMat Waste Water, EU-L-2 as CRM and result was found to be in good agreement with the certified value. The suggested method can be used for selenium species of selenite, selenate, and total organic selenium at μg/L level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.