Abstract

Ion chromatography (IC) coupled with inductively coupled plasma mass spectrometry (ICP–MS) was systematically investigated for determining the speciation of chromium in environmental samples. Firstly, the stability of complexes formed by Cr(III) with various aminopolycarboxylic acids was studied by electrospray ionization mass spectrometry (ESI–MS). The results showed that [Cr(EDTA)] − was stable in solution. Secondly, various mobile phases were examined to separate Cl − from chromium species by IC to avoid Cl − interference. The separation of [Cr(EDTA)] − and Cr(VI) was achieved on a new anion-exchange column (G3154A/102) using a mobile phase containing 20 mM NH 4NO 3 and 10 mM NH 4H 2PO 4 at pH 7.0 without Cl − interference. Detection limits for chromium species were below 0.2 μg/L with a direct injection of sample and without prior removal of interferences from the matrix. Finally, the proposed method was used for the determination of chromium species in contaminated waters.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call