Abstract

A procedure for the determination of selenium species in seawater is proposed. The speciation scheme is based on different sample treatments followed by the quantification of Se(IV) by differential pulse cathodic stripping voltammetry. To remove the organic matter present in seawater which interferes in the electrochemical step a XAD-2 resin is used. So the selenite concentration can be directly determined. After passage through an IRA 400, selenide is collected in the percolated solution. Following conversion into the tetravalent state, selenium is determined. UV photolysis at basic pH is used to convert all the selenium species to Se(IV) which is the electroactive species. The procedure is applied to seawater samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.