A specialized isotope mass spectrometer for determining the isotopic ratio of lithium by ERIAD (electrospray with atomization in an ion source) is described. The device uses a Mattuha-Herzog-type mass analyzer with double focusing to use ion beams with a large spread in energy and simultaneous registration of signals of both lithium isotopes (6Li and 7Li), and a two-channel receiver based on a doublet of microchannel plates. The gas-dynamic interface is built according to the Kontorowitz-Gray scheme; from the gas-dynamic point of view, it is "long", that is, the distance between the nozzle and the skimmer is 4 times the size of the "Mach barrel" at the operating pressure (200 Pa). The device is desktop, its weight does not exceed 20 kg. The measurement time is 1000 s, sample replacement time is 100 s, measurement accuracy is ~ 0.1%. Keywords: mass spectrometry, electrospray, ion source, isotopes of lithium.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.