Abstract

In high throughput small-animal positron emission tomography (PET) studies, it can be advantageous to scan more than one animal at the same time. To optimally position these animals in the scanner, both sensitivity and spatial resolution have to be known across the entire field of view (FOV). The Siemens Inveon small-animal PET scanner has been characterized according to NEMA NU 4-2008 standards, but the spatial resolution has only been determined in a small portion of the FOV and only for filtered backprojection (FBP). In this study, the spatial resolution in the entire FOV was determined for the Inveon using FBP, ordered subset expectation maximization in 2 dimensions (OSEM2D) and maximum a posteriori (MAP) reconstructions. OSEM2D produced the best spatial resolution in the axial center of the FOV, but MAP outperformed the two other reconstruction algorithms in terms of uniformity of the resolution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.