Abstract

Computer simulation was performed for the measurement process of the parameters of MIOL-Akkord bifocal diffractive-refractive lenses in which the central area of the diffractive element is reduced and spherical aberrations of the eye are compensated. The spatial-energy characteristics of an ideal diffractive lens are preliminarily calculated using existing formulas. The simulation of the process of controlling the intraocular lens parameters has shown that the intensity distribution along the optical axis does not characterize the diffraction efficiency, which is the total (integrated) intensity in the beam cross section at the focus. It has been found that due to the mutual influence of diffraction orders, it is difficult to measure the absolute diffraction efficiency and it is better to evaluate only the relative efficiency.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.