Abstract

A technique is described for direct and precise monitoring of the spatial dose uniformity of an electrically scanned ion beam. The method employs a two-dimensional 10x10 array of Faraday cups, individually connected to a current integrator, to measure local dose distribution. It also employs a microprocessor for rapid data processing and topographic data display. The technique is applicable to high quality ion implantation processes as a uniformity monitor with an accuracy of 1%. In addition, a new concept for uniform ion beam scanning is proposed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.