Abstract

The spatial distributions of scintillation photons in a plastic scintillation detector were measured using a multi-anode photomultiplier H7546A coupled with 1-mm-diameter optical fibers. A row of several tens of fibers connected to the scintillator generates one-dimensional spatial distributions of photons induced by the swift passage of heavy ions. The pulse heights from each channel change depending on the beam position. This can be utilized to determine the positions of the heavy ions. To test the performance of the proposed detection method, an experiment using a 84Kr beam with intermediate energies ranging from 40 to 85MeV/nucleon was performed at the heavy-ion medical accelerator in Chiba (HIMAC). The photon spatial distributions were successfully observed. By optimizing the photomultiplier bias voltage and threshold in the pulse height analyses, a detection efficiency of 98% and a position resolution of 1.1mm in σ were achieved simultaneously.