Abstract
Purpose: 3D detection of centerpoints of prosthetic cup and head after total hip arthroplasty (THA) using CT. Material and Methods: Two CT examinations, 10 min apart, were obtained from each of 10 patients after THA. Two independent examiners placed landmarks in images of the prosthetic cup and head. All landmarking was repeated after 1 week. Centerpoints were calculated and compared. Results: Within volumes, all measurements of centerpoints of cup and head fell, with a 95% confidence, within one CT-voxel of any other measurement of the same object. Across two volumes, the mean error of distance between center of cup and prosthetic head was 1.4 mm (SD 0.73). Intra- and interobserver 95% accuracy limit was below 2 mm within and below 3 mm across volumes. No difference between intra- and interobserver measurements occurred. A formula for converting finite sets of point landmarks in the radiolucent tread of the cup to a centerpoint was stable. The percent difference of the landmark distances from a calculated spherical surface was within one CT-voxel. This data was normally distributed and not dependent on observer or trial. Conclusion: The true 3D position of the centers of cup and prosthetic head can be detected using CT. Spatial relationship between the components can be analyzed visually and numerically.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.