Abstract

The spatial and spectral resolutions achievable by a prototype rotating prism chromotomosynthetic imaging (CTI) system operating in the visible spectrum are described. The instrument creates hyperspectral imagery by collecting a set of 2D images with each spectrally projected at a different rotation angle of the prism. Mathematical reconstruction techniques that have been well tested in the field of medical physics are used to reconstruct the data to produce the 3D hyperspectral image. The instrument operates with a 100 mm focusing lens in the spectral range of 400-900 nm with a field of view of 71.6 mrad and angular resolution of 0.8-1.6 μrad. The spectral resolution is 0.6 nm at the shortest wavelengths, degrading to over 10 nm at the longest wavelengths. Measurements using a point-like target show that performance is limited by chromatic aberration. The system model is slightly inaccurate due to poor estimation of detector spatial resolution, this is corrected based on results improving model performance. As with traditional dispersion technology, calibration of the transformed wavelength axis is required, though with this technology calibration improves both spectral and spatial resolution. While this prototype does not operate at high speeds, components exist which will allow for CTI systems to generate hyperspectral video imagery at rates greater than 100 Hz.