Abstract

ABSTRACTWe introduced monodisperse calibrant particles into an eight-stage non-viable Andersen cascade impactor (ACI) operated at 28.3 L/min and separately quantified the particle mass captured under each of the four concentric rings of nozzles on stages 0 and 1, the entry and succeeding stages of this impactor. On both stages, we found that each ring of nozzles has a particle capture efficiency behavior that differs from the others, and the fraction of calibrant particles deposited under each of the individual rings of nozzles depended on the particle size. We believe this behavior derives primarily from a radial flow velocity non-uniformity associated with recirculation zones introduced by the 110° expansion angle of the inlet cone. Because of these recirculation zones, the inertia of particles larger than about 5 µm aerodynamic diameter will cause their point-wise local concentration to differ from the concentration at the inlet entry. This concentration maldistribution continues to stage 1 primarily...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.