Abstract

We conducted a feasibility study using a total-variation minimization algorithm for image reconstruction in prospectively gated micro computed tomography (micro-CT). The total-variation (TV) minimization algorithm exploits the sparseness of the image’s gradient magnitude and can successfully reconstruct CT images from undersampled data for which conventional analytic reconstruction algorithms fail. We implemented the algorithm and applied it to sparsely-sampled data for a mouse by using a prospectively gated micro-CT system. The images were successfully reconstructed, and an image similarity index was quantitatively calculated with respect to the reference images reconstructed from fully-sampled data. Compared to a conventional image reconstruction algorithm, the TV-minimization algorithm substantially reduced image inaccuracy related to the image artifacts.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.