Abstract

The environmental scanning electron microscope (ESEM) employs a series of pressure limiting apertures and a differential pumping system to allow for electron imaging at specimen chamber pressures of up to 50 torr. Images rich in secondary electron (SE) contrast can be obtained using the gaseous secondary electron detector (GSED) or ion current (Iion) signals. The GSED and Iion signals are amplified in a gas cascade. SEs emitted from a sample are accelerated through the gas in the specimen chamber by an electric field, EGSED, produced by a positively biased electrode located in the chamber, above the specimen. The accelerated SEs give rise to a cascade ionization process that can amplify the SE signal by up to three orders of magnitude. Electrons produced in the cascade are rapidly swept to the biased electrode and are efficiently removed from the gas. Positive ions produced in the cascade drift away from the electrode with a velocity that is at least three orders of magnitude lower than that of the electrons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.