Abstract

Radial immunodiffusion (RID) is used to quantify IgG concentration in neonatal beef or dairy calf serum; variability has been noted that may affect the precision and accuracy of assay results. We determined the source, range, and homogeneity of variance in the results of a commercial bovine IgG RID assay (Triple J Farm). To estimate the variance in the precipitin ring diameter, we used 6 sera, measured 28 times across 8 plates and 4 lots, and 3 standards with known IgG concentrations, measured 75 times across 69 plates and 5 lots. The source of diameter variance was determined using variance partition coefficients for lot, plate, and repetition. We used 11 different methods to generate standard curves to convert RID precipitin ring diameters to IgG concentrations. The Levene test of homogeneity of variance (α = 0.1) was used to evaluate the equality of variance between the standards or serum precipitin ring diameters and calculated IgG concentrations. Lot and plate contributed minimally to the diameter variance. Precipitin ring diameters had equal variance. Calculated IgG concentrations for serum not requiring dilution had equal variance. A linear equation from aggregated standards, performed within the same day, had greater accuracy for the calculated IgG concentrations of the standards compared to other equation methods. Regardless of standard curve methodology or IgG concentration, variability inherent to the assay limits its clinical usefulness.