Abstract

Radioactive and nonradioactive L-carnitine and acyl-L-carnitine were used to evaluate the washing procedures used during the determination of free, total, short-chain, and long-chain acylcarnitine in human and sheep plasma. The volume of fluid trapped by the protein precipitated by perchloric acid is approximately 24% of the total fluid volume and thus contains 24% of free carnitine and short-chain acylcarnitine. Washing twice with distilled water removes about 25% of the long-chain acylcarnitine along with the trapped free carnitine and short-chain acylcarnitines. Washing the pellet twice with a 60 g/L solution of perchloric acid completely removes the trapped free carnitine and short-chain acylcarnitine but does not remove the bound long-chain acylcarnitines. Thus washing with perchloric acid is essential for accurate measurement of long-chain acylcarnitines in plasma samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.