Abstract

This study investigates photomultiplier tube (PMT) nonlinearities, relevant for lifetime phosphor thermometry, at various decay times to assess and minimize the impact on temperature measurement accuracy. The focus is single-shot measurements performed in harsh environments where phosphor signal attenuation often is a concern. The sensitivity of decay time measurements to changing phosphorescence intensity is therefore investigated. The experimental results show that for the studied phosphors and detectors, shorter decay times between 20 ns and 6 µs, saturation effects in the PMTs decreased the measured decay time with increasing signal attenuation. For longer phosphorescence decay times, in the millisecond regime, nonlinearity effects led to an increase in the measured decay time with increasing signal attenuation. The specific detector nonlinearity response will vary among detectors, but the introduced methodology for detector analysis is a useful resource for assessing and improving accuracy in lifetime phosphor thermometry measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.