Abstract

Objective: Gauge the sound pressure and temperature levels emitted by the equipment of the drying and sterilization rooms of the Material and Sterilization Center and compare them with the current legislation.
 Method: It is an experimental study with a quantitative approach, implemented in a school hospital in Joao Pessoa-PB. The data were obtained by a sound pressure meter and a thermal stress meter in September 2016.
 Results and Discussion: The data evidenced high sound pressure levels, reaching a mean of 71.25 and 76.25 dB (A), in the drying and sterilization rooms respectively. The levels found do not exceed that recommended by NR15 (Regulatory Norm 15) which is 85dB(A), but NBR 10152/87 (Brazilian Norm 10152/87) indicates levels between 45 and 55 dB(A) for acoustic well-being. As for the temperature the measured mean ranged between 24.91 and 25.87 °C.
 Conclusion: The levels of sound pressure and temperatures found in the environment can cause physiopsychological changes in the professionals that works in this services.
 Keywords: Sound pressure; Temperature; Material and Sterilization Center

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.