Abstract

Background: Liver transplantation (LT) remains a potentially haemorrhagic procedure whose perioperative bleeding and transfusion could be better monitored using point-of-care devices. Quantra® is a device based on sonorheometry to assess whole blood clot formation. Our aims were to describe Quantra® parameters during LT and to study their correlations with standard laboratory parameters, and to determine Quantra® cut-off values for thrombocytopenia, hypofibrinogenemia and coagulation factors’ deficit. Methods: In 34 patients undergoing LT, blood samples were collected before surgical incision, 15 min after the beginning of the anhepatic phase, and 15 min after arterial revascularization of the graft. Results: Clotting time (CT) was well correlated with prothrombin (PT) ratio and activated partial thromboplastin time (aPTT) ratio. Platelet contribution to clot stiffness (PCS) was correlated with platelets (ρ = 0.82, p < 0.001) and fibrinogen contribution clot stiffness (FCS) with fibrinogen (Fg) (ρ = 0.74, p < 0.001). CT predicted a PT ratio < 30% with an area under the curve (AUC) of 0.93 (95% CI 0.87–0.98; p < 0.001). PCS predicted a platelet count < 50 G/L with an AUC of 0.87 (95% CI 0.76–0.98, p < 0.001). FCS predicted a Fg < 1.0, 1.2 or 1.5 g/L, with an AUC of 0.86 (95% CI 0.77–094, p < 0.001), 0.82 (95% CI 0.74–0.91, p < 0.001) and 0.88 (95% CI 0.82–0.95, p < 0.001), respectively. Conclusion: Quantra® provides a rapid assessment of haemostasis during LT.