Abstract

Two optimized adaptive digital signal processors were developed to implement the preset time count rate meters. The first processor improves stationary characteristics of the count rate meter by implementing the control of the error of the mean count rate when the changes of the mean count rate are only due to its stochastic nature. The second processor is focused on dynamic characteristics. Its time response is made adaptive to the changes of the mean count rate which are due to increased or decreased radiation levels. The change of the mean count rate is sensed by a specially devised detection algorithm. Three mean count rate meters, based on the developed adaptive digital signal processors, were realized and were used for experimental validation of proposed adaptive digital signal processors. The experimental results, conducted without and with radioisotope for the specified errors of 10% and 5%, showed to agree well with theoretical predictions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.