Abstract
The relation for the sensitivity of the pneumatic detector based on the analysis of the thermal, pneumatic and optical parts of the instrument is derived. From this relation the optimum diameter of the flexible mirror and the optimum gas pressure in the detector cell are calculated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.