Abstract

Dosimetry calculations were made on the basis of recent photon interaction data and compared with existing sources. The build-up effect in muscle, photon attenuation in the source, filtration medium, and muscle, and 12 principal photon lines were incorporated in the calculations, and radium seeds of two standard dimensions were used. The results were within 1% agreement with previous findings for a 1-mCi radium source but showed an apparent discrepancy for a 13.3-mCi radium source filtered through a platinum tube, due to photon attenuation and scatter. These results may be extrapolated to fit circumstances such as drug-impregnated tissue.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.