Abstract

Abstract The Neutriran Albedo Personnel Neutron Dosemeter (NAPND) is based on the combination of a sensitive polymer such as polycarbonate (PC) in contact with 6LiF or 10B pellets in a cadmium cover. By adding a 10B pellet in front, the design of the dosemeter can be such that when worn on the body, direct thermal neutron induced alpha tracks (TNIAT), albedo neutron induced alpha tracks (ANIAT) and fast neutron induced recoil tracks (FNIRT) can be simultaneously detected in a single electrochemically etched PC foil. To establish a national personnel neutron dosimetry service in Iran, different parameters such as the effect of 10B(n,a)7Li convertor thickness. Cd thickness and its diameter, dosemeter distance from the phantom, dosemeter angle with phantom and directional response were studied using different phantoms. Under optimised conditions, a sensitivity of 1500 tracks.cm-2.mSv-1 for 252Cf neutrons with a lowest value of 0.05 mSv was measured. The results of these studies are reported and discussed.