Abstract

Diffusion-ordered NMR spectroscopy (DOSY) allows for accurate molecular weight calibration and determination that can be corrected for solvent influences. Polystyrene and poly(ethylene glycol) standards have been used to calibrate DOSY diffusion data for a variety of solvents, showing a high correlation of data when the bulk viscosity of the solvent is accounted for following the Stokes-Einstein equation. In this way, a type of universal calibration is introduced that allows for determinations of average molecular weight that are at least as accurate as those of traditional size-exclusion chromatography (SEC), if not better. Further, we demonstrate that DOSY calibrations can be used between laboratories, hence removing the need for individual calibration of setups as currently done.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.