Abstract
For X-ray absorption spectroscopy, either in transmission mode with concentrated samples or for dilute samples in fluorescence mode, it is advantageous to improve the signal-to-noise ratio by implementing a slit apparatus. Several investigations into the improvement of measurements when slits and filters are employed have been reported; however, these have always been for a particular design and are not transferable between dissimilar systems. A generalized approach to Soller slit design will be presented which enables a target level of noise rejection to be achieved by varying the number, size and placement of the filter and Soller slit assembly. A procedure for determining the reduction in efficiency of the Soller slits with respect to misalignment with the sample will also be discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.