Abstract
Solid-phase extraction (SPE) was combined with thermal desorption (TD) and gas chromatographic (GC) analysis to determine drugs in urine. The extrattion was performed inside a fritted GC liner using about 5 mg TENAX that was inserted into the liner on top of the frit. After extraction, the liner was placed into the injector of the GC and the analytes were thermally desorbed by using a programmed-temperature vaporiser. Several sorbent materials were investigated for the applicability of SPETD-GC analysis. TENAX proved to be the most suitable sorbent, since hardly any interferences were observed and acceptable absolute recoveries (73 and 74%) were obtained for lidocaine and diazepam. A mass selective detector (MSD) in the selected ion monitoring mode allowed detection of lidocaine and diazepam down to 0.5 ng·mL−1 using 50μL urine. The use of only 5 mg of extraction material allowed rapid extraction, while a 10 m GC column provided a fast chromatographic system. As a results, the total analysis time was less than 20 min, including 5 min for drying the TENAX and 5 min for thermal desorption. Thus, SPETD-GC-MS appears to be a powerful tool for the rapid analysis of biological samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.