Abstract

We used a Q-switched Nd:YAG laser with non-linear optical crystals to produce the 5th (213 nm) and the 4th (266 nm) harmonic frequencies. Using these two wavelengths, we ablated fresh porcine corneas and type I collagen synthetic epikeratoplasty lenticules. For the 213-nm ablation, radiant exposure was 1.3 J/cm2. The ablation rate was 0.23 micron per pulse for the epikeratoplasty lenticules. We examined all tissues with light microscopy, transmission electron microscopy, and scanning electron microscopy. Histology for the 213-nm ablation showed a clean ablation crater with minimal collagen lamellae disruption and a damage zone less than 1 micron. In comparison, the 266 nm radiation showed more charring at the edges with a damage zone approximately 25 microns deep with disruption of the stromal lamella. Our results show that this solid state UV laser is a potential alternative to the excimer laser for cornea surgery.