Abstract

A simple, sensitive, accurate, and selective method for determination of ultratrace levels of Co is modified. The method is based on preconcentration of Co on the PAR-loaded Amberlite XAD-7 at pH 2.0 ± 0.2 for contact time as low as 45 min. The adsorbed cobalt was eluted with concentrated nitric acid and measured by flame atomic absorption spectrometry. Recoveries up to 90% were achieved. The optimized preconcentration method was applied to cobalt determination in natural mineral waters. The detection limit was found to be 0.1 ng mL−1. The relative standard deviation was found to be 13% for 600 mL of 2.0 ng mL−1, for 10 replicate preconcentration procedures. Cobalt concentrations in the studied water samples were found to be in the ranges of 0.5–3.5 ng mL−1.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.