Abstract

Robust and easy-to-handle solid-contact ion-selective electrodes (SC-ISEs) based on graphite paste have been developed for the potentiometric detection of NO-3 and NH+4 in environmental samples. Polypyrrole (PPy) has been used as intermediate layer and solid contact between the electron-conducting graphite paste substrate and the ion-conducting polyvinylchloride (PVC)-containing membrane with the immobilized ionophore. For the nitrate- and ammonium-selective electrodes, tridodecylmethylammonium nitrate (TDMA-NO3) and nonactin have been used as ion-complexing compounds, respectively. Both ISEs show nearly Nernstian response in the linear range from 10-5 mol/L to 10-1 mol/L with average slopes of >58 mV/decade. The ISEs possess short response times (<20 s).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.