Abstract
The general model of solid body calorimeters (uniform solid body of arbitrary form) with one automatically operated adiabatic jacket was considered for the first time. The problem was to understand if there was general relationship between heat and temperature rise when calorimeter had essential temperature gradients and surrounded one automatically operated adiabatic jacket.It was shown that it was linear relation and proportionality coefficient (heat equivalent of calorimeter) did not depend on form of temperature dependence in main period but depend on (i) location of heat source; (ii) thermometer location; (iii) heat exchanging conditions and parameters of regulation. Experimental checking of theoretical considerations showed that under the same heat exchanging conditions and parameters of regulation, the heat equivalent of calorimeter for thermometer located on calorimeter surface differed from heat equivalent of calorimeter for thermometer located at radius r = 0.5 R by 0.2%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.