Abstract

The responses of two silicon on insulator (SOI) 3-D microdosimeters developed by the Centre for Medical Radiation Physics were investigated with a range of different low energy ions, with high linear energy transfer (LET). The two microdosimeters n-SOI and p-SOI were able to measure the LET of different ions including 7Li, 12C, 16O, and 48Ti with ranges below $350~\mu \text{m}$ in silicon. No plasma effects were seen in the SOI microdosimeters when irradiated with the high LET ions. A Monte Carlo simulation using Geant4 was compared to the experimental measurements, whereby some discrepancies were observed for heavier ions at lower energies. This discrepancy can be partly attributed to uncertainties in the thickness of the energy degraders and overlayers of the devices. The microdosimetric measurements of low energy 16O ions were obtained and compared to a therapeutic 16O ion beam. The radiation hardness of the two devices was studied using the ion beam induced charge collection technique. Both types of the microdosimeters when biased had no essential changes in charge collection efficiency in the sensitive volume after irradiation with low energy ions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call