Abstract

In this short communication we describe some software techniques for electron energy-loss spectrum measurement. We prepared DigitalMicrograph (Gatan) scripts for multiple spectrum acquisitions, quasi-simultaneous acquisition of low-loss and core-loss spectra, energy drift correction, and other operations. Narrow zero-loss spread of 0.27 eV is demonstrated using a 300 kV field-emission transmission electron microscope (TEM) (Hitachi, HF-3000) and a post-column energy filter (Gatan, GIF2002). The core-loss spectrum is acquired with an energy resolution of 0.36 eV with high reproducibility. The present software techniques effectively achieve the intrinsic energy resolution of electron sources. Sample scripts are provided in the Appendix.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.