Abstract
A newly developed software-based compensation of geometrical misalignments is applied to several computed tomography (CT) instruments. The actual geometry of each CT instrument is measured using a dedicated procedure. Then, given the measured geometries, software-based compensation is applied. Dimensional measurements on a validation object show a consistent reduction in observed errors as a result of the compensation. Robustness of the compensation method is confirmed also on a CT instrument with severe misalignments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.