Abstract

Soft tissue movement between reflective skin markers and underlying bone induces errors in gait analysis. These errors are known as soft tissue artifact (STA). Prior studies have not examined how STA affects hip joint angles and range of motion (ROM) during dynamic activities. Herein, we: 1) measured STA of skin markers on the pelvis and thigh during walking, hip abduction and hip rotation, 2) quantified errors in tracking the thigh, pelvis and hip joint angles/ROM, and 3) determined whether model constraints on hip joint degrees of freedom mitigated errors. Eleven asymptomatic young adults were imaged simultaneously with retroreflective skin markers (SM) and dual fluoroscopy (DF), an X-ray technique with sub-millimeter and sub-degree accuracy. STA, defined as the range of SM positions in the DF-measured bone anatomical frame, varied based on marker location, activity and subject. Considering all skin markers and activities, mean STA ranged from 0.3 cm to 5.4 cm. STA caused the hip joint angle tracked with SM to be 1.9° more extended, 0.6° more adducted, and 5.8° more internally rotated than the hip tracked with DF. ROM was reduced for SM measurements relative to DF, with the largest difference of 21.8° about the internal-external axis during hip rotation. Constraining the model did not consistently reduce angle errors. Our results indicate STA causes substantial errors, particularly for markers tracking the femur and during hip internal-external rotation. This study establishes the need for future research to develop methods minimizing STA of markers on the thigh and pelvis.