In this study, a prototype of an inertial sensor-based gait sonification system was tested for the purpose of providing real-time gait feedback on the knee angle. The study consisted of two parts: (1) a comparison of the knee angle measurement to a marker-based 3D optical capturing system (Vicon, Oxford, UK) with N = 24 participants and (2) an evaluation four different sonification feedback versions in an accentuation × pitch (2 × 2) design on a sample of N = 28 participants. For the measurement system comparison, the RMSE was 7.6° ± 2.6° for the left and 6.9° ± 3.1° for the right side. Measurement agreement with bias up to -7.5° ± 6.2° (for maximum knee flexion) was indicated by the Bland-Altmann Method. The SPM revealed significant differences between both measurement systems for the area 45-90% (p < 0.001) (left) and the area between 45% and 80% (p = 0.007) (right). For the sonification perception, the variation of pitch had a significant effect on the perception of pleasantness of the sound. No effect was found for the accentuation of the swing or stance phase.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.