Abstract

In the high-temperature method for the determination of sulfur in petroleum products, sodium azide added to the titration cell solution eliminates nitrogen and/or chlorine interferences with the iodometric finish. In a laboratory in which sodium azide is used routinely for this purpose, some laboratory personnel working adjacent to the sulfur analyzer reported irritations such as nasal stuffiness. It was suspected that hydrazoic acid was being released from the titration cell, although the reportedly pungent odor of hydrazoic acid was not evident. Subsequent analysis showed hydrazoic acid being released from the titration cell at the rate of approximately 12 mg/min under the conditions for the automated titration of sulfur dioxide. It was estimated that hydrazoic acid vapors at a concentration as low as 0.5 ppm can cause some discomfort to laboratory personnel. The titration cell was subsequently placed within an exhaust hood.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.