Abstract

Management of socket fit is challenging for people using lower-limb prostheses because of residual limb volume fluctuation throughout the day. Releasing socket pressures during sitting (partial doffing) may help users increase their limb volume after they have undergone volume loss earlier in the day. The purpose of this research was to develop and evaluate a system to allow for quick and easy locking pin and socket panel release during sitting and relock upon standing. The system was to allow the partial doff tether length to be custom set for each user, accomplish release and relock in less than 2.0s each, require only one hand, and require a finger push force comparable to a push button on a phone. A motor-driven release/relock system (<240g build weight) housed within the socket adjusts locking pin tether length, and an instrumented ratcheting dial adjusts socket panel position. Three participants with a trans-tibial amputation operated the system properly using one hand. For a partial doff, users preferred a tether length between 5 and 6cm. All users executed release within 1.5s and relock within 1.5s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.