Abstract
Abstract A ridge carbon beam filter for Spreading Out the Bragg Peak (SOBP) up to 20 mm in the targeted area has been designed and experimentally tested. The instrument was constructed of Plexiglas with a density of 1.165 g/cm3 and has a “teeth” structure with a period of 2.5 mm. The design is based on a simple analytical model and detailed GEANT4 Monte Carlo simulations. In-beam test measurements were performed with a 500-MeV/u 12C beam at the Nuclotron facility of the Joint Institute for Nuclear Research (JINR) in Dubna. The filter can be used as a static device positioned in front of patients undergoing carbon therapy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.