Abstract

We present and analyze a technique for snapshot imaging spectropolarimetry. The technique involves the combination of channeled spectropolarimetry with computed tomography imaging spectrometry (CTIS). Channeled spectropolarimetry uses sideband modulation to encode the spectral dependence of all four Stokes parameters in a single spectrum. CTIS is a snapshot imaging spectrometry method in which a computer-generated holographic disperser is employed to acquire dispersed images of the target scene, and both spatial and spectral information is reconstructed using the mathematics of computed tomography. The combination of these techniques provides the basis for a snapshot imaging complete Stokes spectropolarimeter that can be implemented with no moving parts. We review design considerations for the spectropolarimeter and present preliminary simulation results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.