Abstract

Real-time detection of L-lactate is crucial in monitoring tissue oxygenation and organ metabolism in surgery patients during the perioperative period. Traditional commercial L-lactate detection techniques using bulky and expensive instruments are time-consuming, hindering timely feedback in the operating room. Herein, utilizing the modified screen-printed carbon electrode, a portable and user-friendly smartphone-based electrochemical biosensing system, integrated with a miniaturized potentiostat via wireless transmission, was successfully constructed for bedside detection of L-lactate. It has a wide sensing range from 0.05–10 mM and a low limit of detection (9.1 μM). Bland–Altman analysis shows an acceptable agreement with the values of a mean difference (0.114 ± 0.3482) between results obtained with the commercial blood gas analysis instrument and the developed portable system. Accordingly, the developed portable system exhibits a great potential in point-of-care testing owing to its portability and rapid response in L-lactate detection during perioperative monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.