Abstract

Linezolid (LIN) is considered the fi rst available oxazolidinone antibacterial agent. It is susceptible to hydrolysis and oxidation. Five simple, accurate, sensitive and validated UV spectrophotometric methods were developed for LIN determination in the presence of its alkaline (ALK) and oxidative (OXD) degradation products in bulk powder and pharmaceutical formulation. Method A is a second derivative one (D 2 ) in which LIN is determined at 240.9 nm. Method B is a pH-induced differential derivative one where LIN is determined using the fourth derivative (D 4 ) of the difference spectra (ΔA) at 285.3 nm. Methods C, D, and E are manipulating ratio spectra, where C is the double divisor-ratio difference spectrophotometric one (DD-RD) in which LIN was determined by calculating the amplitude difference at 243.7 and 267.6 nm of the ratio spectra. Method D is the double divisor-fi rst derivative of ratio spectra (DD-DD 1 ) in which LIN was determined at 270.2 nm. Method E is a mean centering of ratio spectra one (MCR) in which LIN was determined at 318.0 nm. The developed methods have been validated according to ICH guidelines. The results were statistically compared to that of a reported HPLC method and there was no signifi cant difference regarding both accuracy and precision.